HI9829 Multiparameter pH/ISE/EC/DO/Turbidity Waterproof Meter with optional GPS

fisher brand

The HI9829 is a waterproof portable logging multiparameter meter that monitors up to 14 different water quality parameters.

The microprocessor based multi-sensor probe allows for the measurement of key parameters including pH, ORP, conductivity, dissolved oxygen, turbidity, ammonium, chloride, nitrate, and temperature. The probe transmits readings digitally with options to log data while disconnected from the meter. An optional GPS provides location tracking of measurements. The complete system is simple to setup and easy to use. The HI9829 is highly customizable and supplied with all necessary accessories, packaged in a durable carrying case.

Save on calibration solutions when you bundle!

  • Optional GPS provides location tracking of measurements.
  • Color coded, field replaceable sensors
  • Great for environmental, field, and laboratory testing.
Category:
Request a Quote

Description

Built in GPS Receiver

  • 12 channel GPS receiver and antenna
  • Tracks location using satellites to within 30 feet
  • Latitude and longitude recorded with readings
  • Location naming for GPS coordinates
  • Distance from current location to stored location
  • Hi929829 multiparameter water quality meter software supplied integrates data collected with Google maps
    • Measurement readings including GPS coordinates
    • Date and time
    • Named locations and remarks
    • Sample number

Google maps is a registered trademark of Google, Inc. HANNA Instruments® has no affiliation with Google.

Two probes to choose from (Basic or Logging)

*select option at checkout

The HI7609829 (basic) and HI7629829 (logging) are multiparameter probes for use with the HI9829 portable meter.  It is an option to choose which probe will be supplied with the HI9829. By default the HI9829 and corresponding probe will be supplied with a pH/ORP, conductivity and dissolved oxygen sensors. Either probe can be upgraded to measure turbidity with  a turbidity/conductivity sensor.

HI7609829 Basic Probe

  • Digital probe that is less than 2 inches in diameter
  • 3 color-coded sensor inputs
  • IP68 rating for continuous immersion in water
  • Multistrand –mulitconductor cable rated for 150 lbs (intermittent use)
  • Available with 4, 10, or 20 meter cable
  • Weighted shield protects all sensors.
  • HI7629829 Autonomously Logging Probe

    • Same as basic probe but with logging capability
    • Logs data independent of HI9829 meter
    • Operates on 4 “AA” batteries for up to 70 days*
    • Stores up to 140,000 single parameter readings or 35,000 for all parameters
    • Logs data from 1 second to 3 hours
    • Choose to be supplied with EC or EC/Turbidity sensor
    • Supplied with USB cable for direct connection to Window’s compatible PC

    *All channels logging at 10 minute intervals with no averaging.

  • Waterproof Protection – The water quality meter is enclosed in an IP67 rated waterproof casing and can withstand immersion in water at a depth of 1 m for up to 30 minutes. The probe features an IP68 rating for continuous immersion in water.

    Backlit Dot Matrix LCD Display – The HI9829 features a backlit graphic LCD with on-screen help and the capability to display up to twelve parameters simultaneously. The graphic display allows for the use of virtual keys to provide for an intuitive user interface.

    Intuitive Keypad – The fitted rubber keypad has dedicated keys for power, backlight, up/down arrows, help and alphanumeric characters. The meter also features two virtual soft keys that navigate the user through the configuration of each parameter, meter setup, and logging of data. The interface is intuitive for any user’s level of experience.

    Dedicated Help Key – Contextual help is always available through a dedicated “HELP” key. Clear tutorial messages and directions are available on-screen to quickly and easily guide users through setup and calibration.  The help information displayed is relative to the setting/option being viewed.

    Auto-sensor Recognition – The probe and meter automatically recognize the sensors that are connected. Any ports not used on the probe will not have the parameter displayed or be configurable.

    Automatic temperature compensation – Integrated temperature sensor allows for automatic temperature compensation of pH, conductivity, and dissolved oxygen measurements.

    Automatic Barometric Pressure Compensation – The meter features a built-in barometer with user-selectable units for dissolved oxygen pressure compensation.

    Quick Calibration – Quick Calibration provides a speedy, single point calibration for pH, conductivity, and dissolved oxygen. Standard calibration options are available including pH up to three points, Conductivity at one point and dissolved oxygen up to two points.

    GLP Data – HI9829 includes a GLP feature thatallows users to view calibration data and calibration expiration information at the touch of a key. Calibration data includes date, time, buffers/standards used for calibration, and slope characteristics.

    Data Logging – The HI9829 allows users to store up to 44,000 continuous or log-on-demand  samples with logging intervals from one second to three hours.

    Graphing Capability – Trend graphing with sample date and time stamp may be viewed on the display or transferred to a pc.

    PC Connectivity – Logged data can be transferred to a Windows compatible PC with the included HI7698291 USB adapter and HI929829 software.

    Long Battery Life – The display of the meter has a battery icon indicator to show the remaining power. The meter is supplied with four 1.5V “C “ NiMH rechargeable batteries that provide up to 140 hours of battery life*

    Rugged thermoformed carrying case – The HI9829 meter, probe, and all accessories are supplied in a rugged carrying case designed to provide years of use. The inside compartment of the carrying case is thermoformed to securely hold and protect all of the components.

Specifications

pH Range 0.00 to 14.00 pH
pH Resolution 0.01 pH
pH Accuracy (@25ºC/77ºF) ±0.02 pH
pH Calibration automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
mV Range ±600.0 mV
mV Resolution 0.1 mV
mV Accuracy ±0.5 mV
ORP Range ±2000 mV
ORP Resolution 0.1 mV
ORP Accuracy ±1.0 mV
ORP Calibration automatic at one custom point (relative mV)
EC Range 0 to 200 mS/cm (absolute EC up to 400 mS/cm)
EC Resolution Manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm, Automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm, Automatic (mS/cm): 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm
EC Accuracy ±1% of reading or ±1 µS/cm whichever is greater
EC Calibration automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
TDS Range 0 to 400000 ppm (mg/L); (the maximum value depends on the TDS factor)
TDS Resolution Manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L), Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L), Automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
TDS Accuracy ±1% of reading or ±1 ppm (mg/L), whichever is greater
TDS Calibration based on conductivity or salinity calibration
Resistivity Range 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm
Resistivity Resolution dependent on resistivity reading
Resistivity Calibration based on conductivity calibration
Salinity Range 0.00 to 70.00 PSU
Salinity Resolution 0.01 PSU
Salinity Accuracy ±2% of reading or ±0.01 PSU whichever is greater
Salinity Calibration based on conductivity calibration
Seawater σ Range 0.0 to 50.0 σt, σ₀, σ₁₅
Seawater σ Resolution 0.1 σt, σ₀, σ₁₅
Seawater σ Accuracy ±1 σt, σ₀, σ₁₅
Seawater σ Calibration based on conductivity or salinity calibration
Ammonium-Nitrogen Range 0.02 to 200.0 ppm Am (as NH4 +-N)
Ammonium-Nitrogen Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200.0 ppm
Ammonium-Nitrogen Accuracy ±5 % reading or 2 ppm
Ammonium-Nitrogen Calibration 1 or 2 point, 10 ppm and 100 ppm
Chloride Range 0.6 to 200.0 ppm Cl (as Cl- )
Chloride Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200.0 ppm
Chloride Accuracy ±5 % of reading or 2 ppm
Chloride Calibration 1 or 2 point, 10 ppm and 100 ppm
Nitrate-Nitrogen Range 0.62 to 200.0 ppm Ni (as NO3 – -N)
Nitrate-Nitrogen Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200 ppm
Nitrate-Nitrogen Accuracy ±5 % of reading or 2 ppm
Nitrate-Nitrogen Calibration 1 or 2 point, 10 ppm and 100 ppm
Turbidity Range 0.0 to 99.9 FNU; 100 to 1000 FNU
Turbidity Resolution 0.1 FNU from 0.0 to 99.9 FNU 1 FNU from 100 to 1000 FNU
Turbidity Accuracy ±0.3 FNU or ±2 % of reading, whichever is greater
Turbidity Calibration Automatic 1, 2, or, 3 points at 0, 20 and 200 FNU, or custom
DO Range 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L)
DO Resolution 0.1%; 0.01 ppm (mg/L)
DO Accuracy (@25ºC/77ºF) 0.0 to 300.0% saturation: ±1.5% of reading or ±1.0% saturation whichever is greater, 0.00 to 30.00 ppm (mg/L): ±1.5% of reading or ±0.10 ppm (mg/L), whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ±3% of reading, 300.0 to 500.0% saturation: ±3% of reading
DO Calibration automatic one or two points at 0 and 100% or one custom point
Atmospheric Pressure Range 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
Atmospheric Pressure Resolution 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
Atmospheric Pressure Accuracy ±3 mm Hg within ±15°C from the temperature during calibration
Atmospheric Pressure Calibration automatic at one custom point
Temperature Range -5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K
Temperature Resolution 0.01 K, 0.01 °C, 0.01 °F
Temperature Accuracy ±0.15 °C; ±0.27 °F; ±0.15 K
Temperature Calibration automatic at one custom point
Temperature Compensation automatic from -5 to 55 °C (23 to 131 °F)
Logging Memory 44,000 records (continuous logging or log-on-demand of all parameters)
Logging Interval one second to three hours
PC Connectivity USB (with HI 929829 software)
Battery Type/Life 1.2V (4) NiMH, rechargeable batteries, size C type; up to 140 hours depending on configuration
Waterproof Protection IP67
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 221 x 115 x 55 mm (8.7 x 4.5 x 2.2”)
Weight 750 g (26.5 oz.)
GPS 12 channel receiver 10 m (30 ft) accuracy
Ordering Information The HI9829 and HI98290 (GPS model) are supplied with a HI710140 hard carrying case, HI7612VDC adapter, HI7698292 probe maintenance kit, HI929829 PC application software, HI7698291 USB cable (PC to meter), HI710045 power supply cable, HI710046 cigarette lighter power cable, HI9828-25 quick calibration solution, HI920005 iButton with holder (5), HI7609829-1 pH/ORP sensor, HI7609829-2 galvanic DO sensor, HI7609829-3 EC sensor, HI7698290 short calibration beaker, HI7698295 short protective weighted shield, NiMH rechargeable “C” cell batteries (4), 12VDC power adapter, and instruction manual. 

Probes selected with turbidity option will include the HI7609829-4 EC/turbidity sensor, HI7698293 long calibration beaker and HI7698296 long protective weighted shield in place of the EC sensor, short calibration beaker and short protective shield. Probes selected with turbidity option will also include turbidity calibration solutions (230 ml); HI9829-16 0 FNU, HI769829-17 20 FNU, and HI9829-18 200 FNU.

Warranty 2 years (probe 1 year, sensor 6 months)

Reviews

There are no reviews yet.

Be the first to review “HI9829 Multiparameter pH/ISE/EC/DO/Turbidity Waterproof Meter with optional GPS”

Product Enquiry

Name
Email
Phone
Enquiry